8 Troubleshooting
75 R-220 Pro Continuous Operation Manual, Version A
Other error situations
In case the temperature reading at the display for vapor or cooling water is missing, check the respec-
tive sensor cable and plug. Exchange a defective sensor and call service if the problem is persistent.
8.3 Resettable automatic system fuses
Fuse triggered
Fuse ok
F1 — F4 (12A, 250V)
Resetting a system fuse
Switch o the Rotapavor
®
Push in the fuse for reset
Switch on the Rotavapor
®
NOTE
Device fuses may be triggered occasionally due to voltage peaks under high system load. When fuses
trigger frequently, inform your customer service! Do not try to repair the fuses!