5 Putting into operation
48 R-220 Pro Continuous Operation Manual, Version A
5.13 Heating bath setup for 4.2 kW heating element
The heating element holder is labelled with the heating power data. For the 4.2 version only water
must be used as heating transfer medium. Because of the enormous heat input there is a risk of
smoke formation and re when using other heating transfer media than water!
How to set the maximum setpoint temperature
Limiting the maximal temperature set-point to a value below 100 °C can help to eliminate processing
errors, e.g. when working with heat-sensitive sensitive substances.
AERATE
Temp
+
Set Temp
SET
Store setting
X Keep the "SET" button pressed while switching on
the Rotavapor® to enter the setup menu.
X Set the temperature limit via the turning knob. The
set value is displayed as bath temperature value
on the display.
X Press "AERATE" button to store the setting and to
leave the menu.