4 Description of function
29 R-220 Pro Continuous Operation Manual, Version A
Position Drawing Description
Vacubox. For a detailed description see I-300 Pro operation
manual.
See master drawing for appearance
and mounting position.
Aerate valve (parallel to controller aeration valve)
Fuse triggered
Fuse ok
Resettable fuses. Gently push them back into prestressed posi-
tion when fuses are activated.
Mains input cable to supply the Rotavapor
®
. Output socket to
supply a vacuum pump. Mind the voltage and maximal output
current, when installing a vacuum pump! The combination
of Rotavapor
®
and vacuum pump must not violate the local
electrical safety regulations. Contact an electrician before
connecting any device.