8 Troubleshooting
71 R-220 Pro Continuous Operation Manual, Version A
Errors and their remedy
Error Additional error indicator on display Possible cause Remedy
E01 Heating bath icon ashes. Bath temp. sensor not connected or
defective.
Switch o the unit. Wait for the bath
to cool down a little before restart.
Call service if problem is persistent.
E02 Heating bath icon ashes. The bath does not reach the
intended maximum height.
Lift motor is defective or
blocked.
Press the Aerate button to conrm
the error message.
Switch o the unit. Check if lift is
blocked.
Call service if problem is persistent.
E03 Drive unit icon ashes. The rotar y drive does not reach the
set sppec value.
Drive defective or blocked.
Press the Aerate button to conrm
the error message.
Switch o the unit and check if
drive is blocked, rotate evaporating
ask manually.
Call service if problem is persistent.
E04 n.a. Battery voltage too low.
PB battery weak or defective.
Press the Aerate button to conrm
the error message.
Switch on the system over night to
allow recharging of the battery.
Call service if problem is persistent.
E05 n.a. Hardware malfunction. Restart the unit.
Call service if problem is persistent.
E06 n.a. Shortcut at aeration valve. Unplug valve and restar t system for
causal research.
Exchange valve if defective or run
evaporation with manual aeration.
Call service if problem is persistent.
E07 n.a. Shortcut at cooling water valve. Unplug valve and restar t system for
causal research.
Exchange valve if defective or run
evaporation without cooling water
valve.
Call service if problem is persistent.