Inspection report
(normalised to
rated value of
mains voltage)
Result: Bed prot.
class II (type BF)
0.1 mA mA
In case of measured voltage external
conductor - earth
volt:
Inspection report
III. Functional check OK Not
OK
Description of
defect
Functional check of the electrical components
What? How?
End of travel cut-out of the motors Automatic cut-out
External power supply / handset No ‘rattling’ when shaken?
Handset: Operating function, lock-
ing function
Perform the test acc. to
Part C: Handset » 8
Motors Abnormal noise develop-
ment (rattling, uneven run-
ning)
Strain Relief of mains cable (if
mains cable available)
Mains cable firmly fas-
tened
Functional check of the mechanical components
Joints and pivots Smooth operation
Grab handle with strap Securely fixed when load
tested under approx. 75
kg load (hang from it brief-
ly with two hands)
Castors Moving and braking
Emergency release of the backrest Test according to instruc-
tion manual
Safety sides Securely engaged, secure
position, unlocking
Maintenance
Part B: Operator and Technical Personnel 53