7.6 Initialising the control unit
The initialisation procedure must be carried out by technical personnel only.
When initialising the control unit, it is important that none of the functions are locked on
the handset!
Any restrictions on lowering the mattress base height which have been saved will be lost.
There must be no pauses longer than 6 seconds between the steps without pressing a
key. The system will otherwise switch back to the normal operating mode.
5. Remove the old control unit from its bracket and fit the new control unit in the same
way.
6. Re-insert all plug connections in the corresponding sockets. Make sure that the O-rings
on the plugs are present and undamaged. These seal the plugs into the control unit.
7. When routing the handset cable, ensure that it cannot be damaged by any moving
parts of the bed.
8. Put the plug cover back in place. This prevents all the plugs from being pulled out of
the control unit.
9. Perform a new initialisation of the control unit; see Part B: Initialising the control
unit » 60.
10.
Check the motorised bed adjustment functions; see Part C: Operation » 8!
This is necessary after installing a new control unit, or when replacing one or both lift-
ing drives of the mattress base height adjustment, or if height adjustment of the mat-
tress base is not possible or only partly possible.
This involves assigning the path impulses of the lifting drives to a reference point by means
of a reference run.