6.2.2 Inspection report
Limit value:
Leakage current I Ablsmaller than 0.1 mA.
Inspection cycle:
We recommend an annual inspection and functional check. If this test is passed, electrical
measurement every ten years is sufficient.
In the case of verifiable compliance with 2% error rate (see also DGUV regulation 3: §5, Ta-
ble 1B), the inspection cycle of the inspection and function test can be extended to a maxi-
mum of 2 years.
The inspection report templates shown on the pages that follow should be used.
The following is an inspection report template for inspecting electro-medical equipment in ac-
cordance with DIN EN 62353 (latest issue):
Inspection report
Customer / Medical facility / Practice:
Address:
Carried out: [ ] Repeat inspection [ ] Inspection prior to initial operation (reference value)
[ ] Inspection following repairs/maintenance
Equipment type: [ ] Hospital bed [x]
Care bed
Protection class: [ ] I [x] II
Bed type: Regia Inventory number:
Location: Serial number:
Application environment
(IEC60601-2-52): [ ] 1 [ ] 2 [x] 3 [x] 4
[ ] 5
Manufacturer: Burmeier GmbH & Co.
KG
Applied parts: Mattress base, headboard, footboard, safety
sides
Testing equipment used (type/inven-
tory no.):
1:
Medical Device Regulation classifica-
tion: Class I
2:
Maintenance
50 Part B: Operator and Technical Personnel