Using Luna
Right Mainstem Intubation
If a learner performs an intubation improperly that results in a right mainstem intubation, the left
side of the chest will not rise when ventilated and sounds in the left lung will be absent upon
auscultation. Additionally, an indicator appears on the Airway avatar in the Maestro software and an
event for right mainstem intubation will be captured in the SCE Event log.
Laryngospasm
The Luna simulator comes with a manual laryngospasm feature to increase the difficulty of
intubation.
To activate the laryngospasm feature:
1. Enable Laryngospasms in the Maestro software. Tap the Respiratory icon on the SCE run
screen.
2. Slide the Laryngospasm parameter toggle to the ON position.
3. Connect the IV tubing to the Laryngospasm port on the back of the manikin’s neck.
4. Attach a 50 ml syringe filled with air to the tubing and instill 35 - 40 ml of air into the port
to close the cords.
5.Detach syringe and tubing from the Laryngospasm port.
To resolve the laryngospasm feature, slide the Laryngospasm parameter toggle to the OFF position.
Bag Valve Mask (BVM)
The manikin can receive airway management interventions using a bag valve mask (BVM).
To perform this intervention:
1. Adjust infant’s head to sniffing position.
2. Place the BVM over the manikin’s nose and mouth.