EasyManua.ls Logo

Cannon CAV 4.2 - Train sensors (CAV 4.2 GUI)

Cannon CAV 4.2
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAV® 4.2 Operator's Manual, Rev. 2.0
22
Getting Started CANNON®
Figure 18 — Train screen
Train sensors (CAV 4.2 GUI)
Train the sensors whenever a CAV® 4.2 bath is to operate at a new temperature. Once trained,
the calibration values for a particular temperature are saved and will be recalled if the CAV®
4.2 is set to run at that temperature again. Retraining for a particular temperature is only
necessary when the viscometer tube is changed or if sensor performance changes over time.
Table 14 — Tools and materials required
Description
Description
Assembled CAV® 4.2
Sample test oil
Sample vials
Sample preparation materials
Notes: The ubbelohde tubes in each bath may have different viscosity
limits.
Temperature calibration must be performed first.
1. Insert an empty vial into position 1 of the carousel for the bath to be trained.
2. Fill a vial with a test solution that has a viscosity approximately 3 times higher than the
lower viscosity limit of the ubbelohde tube. Insert the vial into position 2.
3. From the main screen, click the tube button ( ) for the appropriate bath. The train
screen opens.
Note: Do not check the
Prewash or Trip Points
Only boxes.
4. Click Start Training.
5. When training is complete, repeat
steps 1 through 4 for the second bath.

Table of Contents