CAV® 4.2 Operator's Manual, Rev. 2.0
43
CANNON® Operating the CAV 4.2
Connect VISCPRO II to the CAV® 4.2
VISCPRO II and the CAV® 4.2 must be set up and communicating prior to attempting to
establish a connection. Refer to Connecting with VISCPRO II for instructions.
1. Start VISCPRO II on the PC.
2. On the CAV 4.2 GUI, touch the Local Mode/Remote Mode button to toggle the
instrument state. The icon will turn green to indicate that the CAV 4.2 has found
VISCPRO II.
3. In the VISCPRO II Menu, click Main > Poll for Instruments. If an instrument is located, an
associated message box will be added to the status bar on the bottom of the VISCPRO II
window.
Note: CAV 4.2 are grouped according to the Ethernet port number on
which they communicate. Since VISCPRO II can only communicate on
one Ethernet port at a time, there will be only one group of CAV 4.2
instruments.
4. Click Main > View Instrument and select the CAV 4.2 instrument group. The CAV 4.2
Instrument Window opens.
Caution: When multiple instruments (CAV 2100/2200, miniAV, CAV
4.2) are connected to the same instance of VISCPRO II, verify that all
instruments have completed their current operations before closing
VISCPRO II.
Set bath temperature (VISCPRO II)
1. In the VISCPRO II Menu, select Configure > [instrument group] > [instrument] > Tray
Settings: Temperature.
2. Adjust the Temperature Units, Target Temperature, and Temperature Tolerance as
necessary.
3. Click Save; wait for the system to save the changes, and then click Close.
4. Wait for the bath to adjust to the new temperature.