EasyManua.ls Logo

Cannon CAV 4.2 - Table 17 - VISCPRO II Registry Entries

Cannon CAV 4.2
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAV® 4.2 Operator's Manual, Rev. 2.0
27
CANNON® Connecting with VISCPRO II
The following registry keys are set to default values during installation of VISCPRO. You
cannot edit them from within the VISCPRO software. The only way to change these values is
by directly editing the registry.
Caution: Editing registry entries may be hazardous to the health of
your Windows® installation.
Table 17 — VISCPRO II registry entries
Registry Key
Default value
Description
EthernetPortNumber
1982
TCP port on which VISCPRO listens
EthernetServerIPAddress
10.0.1.2
IP address of the Ethernet connection
that VISCPRO uses
MaxAllowedInstruments
4
Maximum number of instrument to
which VISCPRO will connect. VISCPRO
polls for instruments in sequential order
of the InstrumentAddress. Once the
maximum number of allowed instrument
has been reached, it stops polling.
MaxPollResponseWaitTime
2000
(2 seconds)
Maximum time in milliseconds that
VISCPRO waits for an instrument
response when polling.
MaxResponseWaitTime
20000
(20 seconds)
Maximum time in milliseconds that
VISCPRO waits for an instrument
response.

Table of Contents