5 Maintenance
5.1 Recommended actions and precautions
The analyzer is easy to use and it is normally free of maintenance.
Do not spill any liquid over the analyzer. I case of spilled liquid suck it immediately up
with a suitable tissue or use a pipette to remove it out of cells.
Wipe the surface with a disinfection solution made for laboratory instruments once a
week for your own safety.
For measuring safety it recommended to use frequently a suitable precision or accura-
cy control material.
Problem cases
MLB242 will not turn on obviously there is a problem in the power supply. Check the
correct connection of the external power supply between the appliance and the electric-
ity network. Ask a technician to check the operating voltage.
Measuring cells are contaminated by a liquid.
Switch o󰀨 the device immediately!
Remove the uid such as using a pipette, siphon etc. or in the case of larger quan-
tities of liquid to simply turn the device, allow the liquid to ow out. Then clean the
cell (s) with a lint-free absorbent felt to remove liquid remains. In the case of biolog-
ical materials, clean the cell (s) with an appropriate disinfectant, see corresponding
notes in the chapter! Safety instructions.
Note any error messages when switching on the computer.
If necessary, call the technical service.
Incubation cells are contaminated with uid.
Proceed as described in the preceding case accordingly!
5.2 Software update
The manufacturer will inform about available software updates by E-Mail after a prod-
uct registration and o󰀨er them for download. Proceed as described in Chapter 6.5.