6.4 Technical data
Analyzer Type: Analyzer with photometer technique
Operating system: Linux OS v2.6.33
Main processor: ARM 9 AT91SAM9263
Memories: 128 MB for application routine
2GB SD ship for storing of user data
Measuring head: 4 cells, 2 optics each (dierent dep. on analyzer type)
temp. controlled at 37°C ±1/98.6°F
- automatic calibration and self-testing
- linearity control, if necessary warning message
- drift compensation automatically
- photometric range: 0,0-6,0 Abs
- photometric resolution: A = 0,0001 max
- photometric accuracy: ± 3%
Self-checking: made automatically after power on and after test selection.
Tube sensor: each measuring cell is equipped with a tube insertion Sensor.
Analysis methods: Endpoint, Kinetic and xed time
Incubation block: 37,0°C ±1/98.6°F, temp. controlled, 4 x 4 positions
Cuvettes: Volume at triangular mark: 1000μl
mixed polymer, for optical use
Cups: Polystyrol made
Languages: English, German, Francais, Italiano, Espanol.
Tests selectable: pre-dened test stored in memory
LCD-Display: TFT 4,3” colour with touchscreen function
Connectivity: 1 x USB printer port for servicing and data transfer to host
Bluetooth 2.1 function will be available on later series!
Printer no
System time: real time clock for time and date
Power supply: external AC/DC adapter
Prim. voltage: 100 - 250V, 47 - 63 Hz, 0,8 A
Sec. voltage 24V, 1,4 VA, max.: 3,0 A
Fuses: no
Safety certicates: TÜV/GS, UL, CE
Optional: Lithium battery 5000 mAh will be available on later series!
Environment- rel. humidity: 20% - 80%, no condensing
conditions: temperature range +12 to +30°C / 53.6 to 86°F
Dimensions: height 90 mm, width 145 mm, depth 220 mm
Weight: ca. 765 gr.