8200-M035-O1 Rev C 200 Installation & Technical
25
SETUP AND CALIBRATION, CONT.
A - d (A - d?) - Analog to Digital Filtering
SErSCL (REMOTE INDICATOR SETUP)
If the indicator is to function as a remote indicator, press the ASTERISK/UP ARROW key to
select YES at the SErSCL prompt. If not then press the TARE/ENTER key at the prompt.
IMPORTANT! If the 200 indicator is to be used as the LOCAL and as the
REMOTE, then both indicators MUST have Rev 2.6.3 or greater software installed.
If a combination of a 200 and a 225 indicator are to be used in a LOCAL/REMOTE
configuration, then the remote 200 indicator MUST have Rev 2.6.1 or greater
software installed, and the 225 indicator MUST have Rev 1.0.K or greater software
installed.
LrP= (Local/Remote Port)
NOTE: This will only be displayed if SERSCL was set to YES
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the ASTERISK/UP ARROW key
to toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values
are: 1 for port 1 or 3 for the optional 2xx-RS232 card (if installed).
Refer to the LOCAL/REMOTE SETUP section of this manual for more information and
the interconnect wiring diagram.
dFLt= (Digital Filtering)
With A - d (A - d?) displayed, press the TARE/ENTER key. The display will change to
dFLt=. Press the TARE/ENTER key to show the current value. If the setting displayed is
acceptable, press the TARE/ENTER key to save it. Otherwise, use the ASTERISK/UP ARROW
key to toggle to a new setting and then press the TARE/ENTER key to save it. Allowable
values are: 0, 1, 2 or 3. Note, that if you select 3 (Custom Filtering) two additional prompts will
be displayed.
dFLt=
0
Disabled NO Filtering
1
MINIMAL FILTERING (sample rate = 2)
2
MODERATE FILTERING (sample rate = 1)
3
CUSTOM FILTERING
NOTE: The prompts, F= (Filter Level), h= (High Frequency Filter Value) and b= (Break
Range) will only be displayed if you selected 3 (Custom Filtering) for the dFLt= (Digital
Filtering) prompt.
F= (Filter Level)
Press the TARE/ENTER key to show the current setting for the filter level. The filter level is a
number from 1 to 99 that corresponds to the level of filtering with 99 being the greatest filtering
and 1 the least. To accept the value displayed, press the TARE/ENTER key, otherwise, use the
ASTERISK/UP ARROW key to toggle to a new setting and then press the TARE/ENTER key to
save it.
b= (Break Range)
Press the TARE/ENTER key to show the current setting for the break range. The break
range is a number from 1 to 255 that corresponds to the number of division change to
break out of the filtering. Press the TARE/ENTER key to keep the displayed value or use
the ASTERISK/UP ARROW key to toggle to a new setting and then press the TARE/ENTER
key to save it. Note that selecting 0 disables this feature.