8200-M035-O1 Rev C ļ‚Ÿ 200 Installation & Technical
26
SETUP AND CALIBRATION, CONT.
h= (High Frequency Filter Value)
Press the TARE/ENTER key to show the current setting for the high frequency filter level. The
filter level is a number from 0 to 99 that corresponds to the level of filtering with 99 being the
greatest filtering and 1 the least.
Press the TARE/ENTER key to keep the displayed value or use the ASTERISK/UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Note that selecting 0
disables this feature.
Sr= (Sample Rate)
Press the TARE/ENTER key to show the current setting for the sample rate. The value
displayed is the sample rate in samples per second. Press the TARE/ENTER key to save the
displayed value or use the ASTERISK/UP ARROW key to toggle to a new value (1 to 100) and
then press the TARE/ENTER key to save it.
UnS= (Motion Range)
Press the TARE/ENTER key to view the current setting for the range of motion detection. If the
displayed value is acceptable, press the TARE/ENTER key to save it. Otherwise, use the
ASTERISK/UP ARROW key to toggle to a new range (the number of divisions of change
permitted before indicating unstable) and then press the TARE/ENTER key to save the new
setting. Allowable range values are: 0 through 99 divisions.
SC= (Stable Count)
Press the TARE/ENTER key to view the current setting for the number of consecutive stable
weight readings before indicating stable weight. This helps filter weight readings for stability
when trying to capture stable weight. If the displayed value is acceptable, press the
TARE/ENTER key to save it. Otherwise, use the ASTERISK/UP ARROW key to toggle to a new
value and press the TARE/ENTER key to save the new setting. Allowable values for the stable
count are: 3 through 255.