8200-M035-O1 Rev C ļ 200 Installation & Technical
SETUP AND CALIBRATION, CONT.
Sio (Sio?) - Serial Input/Output
With Sio (Sio?) displayed, press the TARE/ENTER key. The display will change to show the
current setting no. To skip configuring the Sio (serial input/output) and proceed to the
Print? menu, press the TARE/ENTER key again. To configure the Sio, press the
ASTERISK/UP ARROW key (display will change to YES) and then press the TARE/ENTER key.
After pressing the TARE/ENTER key, the display will change to bAUd=.
bAUd= (Serial Port Baud Rate)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the ASTERISK/UP ARROW key to
toggle to a new baud rate for the serial ports and then press the TARE/ENTER key to save it.
Allowable values are:
Prty= (Serial Port Parity)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the ASTERISK/UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values are:
bitS= (Serial Port Data Bits)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the ASTERISK/UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values are: 7
or 8.
StoP= (Serial Port Stop Bits)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the ASTERISK/UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values are: 1
or 2.
Cont1= (Continuous Output Serial Port 1)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the ASTERISK/UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it.
If Cont1= YES (Continuous Output) is selected, an additional prompt, tYPE=
will be displayed.
If Cont1= no (No Continuous Output) is selected, proceed to the Weight On
Demand section.