8200-M035-O1 Rev C 200 Installation & Technical
SETUP AND CALIBRATION, CONT.
tyPE= (Continuous Output Format)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the ASTERISK/UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values are:
8 = SB500 with Traffic Light
SMA – (Type = 0)
If SMA is selected, the data will be transmitted in the following format:
<lf><s><r><n><m><f><xxxxxx.xxx><uuu><cr>
Where:
Z= center of Zero, O = Overcap, E = zero Error,
e = weight not currently being displayed
G = Gross, T = Tare, N = Net
M = Motion, " "(blank) = no motion
Six digits with decimal point
ton, lb , l/o, oz , t , kg , g
SB-400 – (Type = 1)
If SB-400
or Computer is selected, the data will be transmitted in the following format:
<s><xxxxxx><d><uu><m><cc><cr>
Where:
"-" = negative, " " (blank) = positive
Added to string if enabled in setup
tn, lb, l/o, oz, t, kg, g
OC = overcap
CZ = center of zero
MO = motion
ee = weight not currently being displayed
The SB-80, SB-300 (multiple displays not supported) and WinVRS use the SB-400 format.