8200-M035-O1 Rev C 200 Installation & Technical
37
SETUP AND CALIBRATION, CONT.
SB-200 (Type = 2)
If SB-200 is selected, the data will be transmitted in the following format:
<cr><s><xxxxxx><d><c><uu><m> ETX
Where:
cr =
Carriage Return
(hex 0D)
s =
Sign
"-" = negative, " " (blank) = positive
xxxxxx.xxx =
Weight
(with leading zeros)
d =
Decimal point
Embedded into weight (after weight dpp=0)
c =
status
m = motion
o = overcap
e = weight not currently being displayed
uu =
Units
tn, lb, l/o, oz, t, kg, g
m =
Mode
G = Gross, N = Net
ETX =
End of TeXt
(hex 03) MUST terminate ALL serial commands
Rice Lake IQ355 (Type = 3)
If Rice Lake IQ355 is selected, the data will be transmitted in the following format:
<stx><s>xxxxxxx<u><m><s><cr><lf>
AnDFV (Type = 4)
If AnDFV is selected, the data will be transmitted in the following format:
<hdr1>,xxxxx<uu><cr><lf>
WI110 (Type = 5)
If WI110 is selected, the data will be transmitted in the following format:
<m><sp><s>xxxxx<sp><uu><cr><lf>
Number Only (Type = 6)
If number only is selected, the data will be transmitted in the following format:
xxxxxx<cr><lf>
Toledo Short (Type = 7)
If Toledo Short is selected, the data will be transmitted in the following format:
<stx><swa><swb><swc>xxxxxx<cr><sum>
Toledo Long (Type=7)
If Toledo Long is selected, the data will be transmitted in the following format:
<stx><swa><swb><swc>xxxxxxyyyyyy<cr><sum>
Where:
xxxxxx =
Weight
(leading zeros suppressed)
yyyyyy =
Tare Weight
(leading zeros suppressed)