16
3) Fix the calibration weight cover (1) with the
calibration mass cover setscrew (2).
Note
Steps (3) to (5) are only required with the CBW-K
series.
4)Put the built-in calibration weight (3) in place and
tighten the calibration weight stopper screw (4).
5 )Turn both the left and right calibration levers of
the pan to the Measurement position.
6) Place the pan on the balance. For the CBW-K series,
the pan will be easier to place if it is placed slightly
from the rear.
7) The level of the balance is adjusted Note.
Turn the level screws to adjust it so that the air bubble in the level rests inside the red circle.
Check that the balance is steady when adjustment is complete.
Note
Adjusting the level can be done with ease if all 4 level screws are at first grounded equally.
Next, adjust the front and the back using the 2 front feet, and then adjust left and right using
either of the 2 feet on the left and right.
1.2.2.1 Fitting the Vinyl Cover
When using the balance in a location where it is likely to get dirty easily, refer to the diagram
below and affix a vinyl cover by following the procedure below. Remove the pan first if
fitting the vinyl cover after the balance has been assembled.
1) Cut the supplied tape (2) and stick it onto the 4 points on the case (3) (refer to the diagram
below).
2) Fix the vinyl cover (1) onto the double-sided tape (2), ensuring that there are no creases.
3) Fit the pan.
Measurement