2.9.6 Interval Timer
Automatically outputs the displayed value at preset intervals. The symbol is illuminated when the
Interval Timer is activated.
The optional AKB-301 Application Keyboard is convenient when the interval is changed frequently.
1) Using Menu Map item [50], set the output interval (00:01 = 1 sec to 99:59 = 99 minutes 59
seconds).
2) In the interval timer standby state when the
and symbols are both illuminated,
press the
key.
The first data is output. Data is automatically output at the set time intervals.
3) To stop output, press the
key.
The balance returns to the interval timer standby state in step 1).
Notes:
Use the key to erase the tare or zero the balance at any time.
Pressing the key while in the interval timer standby state causes the power supply standby
state to be initiated.
To release the interval timer function, keep pressing the key. This does not reset the
interval timer function to zero.
Using the interval timer function to record data over a long period may cause data error due to
balance drift.
Some instruments receiving the data may not operate normally if the set time interval is short. To
correct this, set the time interval to a longer period. When the set time interval is short and the
instrument connected to DATA I/O is unknown, it is recommended to set the handshake to a
setting other than
[76].
2.9.7 Auto-Memory and Zeroing
Use Menu Map item [51].
This menu is used to weigh a large number of individual samples. The
(Auto-Memory and
Zeroing) symbol is illuminated when this function is active.
1) Load the weighing vessel and press the
key in the Auto-Memory and Zeroing
standby state (
and symbols are lit).
Zeroing occurs.
2) Press the
key.
The
symbol disappears, and Auto-Memory and Zeroing measurement starts.
3) Load the first sample. Each time the stability symbol is lit and the display is a value five times or
more the Zero Range or the
key is pressed, the displayed value is output and zeroing
occurs.