71
is the target value in the Target mode. is displayed for other modes.
*4 This operation is performed only in Target mode.
is displayed for other modes.
*5 Operation of the TARE and PRINT keys are the same as keys on the main balance body.
3.3.2 Example of Use
3.3.2.1 Using with a Comparator
Note: The decimal key of the AKB is not used for setting comparator values. The decimal place is set
automatically according to the display capabilities of the balance.
Preparation
Set the kind of graphic display you desire to use with menu selection.
1) From the mass display, press the key three times and the key once.
2) Use the
and keys to set the desired graphic display mode.
(1) Target display mode
Setting the target value
Example: Press the numeric keys for 10000 and then the [TARGET]key.
The Target value is set to 10000.
Setting the limit value
Example: Press the numeric keys for 5000 and then the [LOWER] key.
The Limit value is set to 5000.
Checking the target value
Press the [TARGET] key to display the set value.
Checking the limit value
Press the [LOWER] key to display the set value.
After checking, press the g/PCS key to return to the mass display.
Note:
is displayed if the [TARGET] key is pressed without being in the Target display
mode.
(2) In Comparator mode (group display and level display)
Setting the upper limit value
Example: Press the numeric keys for 20000 and then the [UPPER] key.
The upper limit value is set to 20000.
Setting the lower limit value
Example: Press the numeric keys for 10000 and then the [LOWER] key.
The lower limit value is set to 10000.
Checking the upper limit value
Press the [UPPER] key to display the set value.