36
6) When correcting the value to zero, press the key.
Press the
key if not correcting it to zero
(although pressing the
key will also leave it uncorrected, this will in
effect mean “Calibration Aborted”).
Note: Changing the
value to zero is equivalent to performing span calibration.
7)
is displayed, indicating the completion of the calibration check.
Note: Examples for interpreting the results of a Calibration Check:
Value Actual Mass
Displayed Mass
-0.0003 30kg 29.9997 (32kg/0.1g balance)
+0.21 400g 400.21 (420g/0.01g balance)
Error codes that may be displayed:
indicates that the value is 1000 counts or more.
indicates that the value is -1000 counts or less.
2.5.4 Span Calibration Using External Weights
1) Verify that the balance is in mass display and unload the sample from the pan.
2) Press the
key once. is displayed.
(If
is not displayed, return to mass display and select Menu Map
item [3].)
3) Press the
key.
The value of the correct calibration mass to be loaded is displayed and blinks.
Changing the Calibration Mass to be Used
Pressing the key allows changes to the weight value. Modify the value
using the
key and key, then press the key. To
interrupt modification, press the
key.
4) Load the indicated calibration weight and press the key.
5) Shortly, zero display blinks. Unload the weight from the pan and press the
key.
is displayed briefly to indicate completion of span calibration.