43
2.8.6 Whole Lighting Mode
Press the key to switch from the standby mode to the mass display. The entire display is
illuminated. This function sets whether the display automatically proceeds to mass display or waits for
a
key command. When the mode has been set to , it is easier to
check the display because the display remains illuminated.
Menu Map item [37]
Display stops at whole lighting. Press the
key to proceed to
the mass display.
Menu Map item [38]
After the whole lighting display has appeared for approximately 0.5 sec, the
display automatically proceeds to the mass display.
2.8.7 Taring/Printing
Determine if the balance should wait for stability before printing when the key is pressed or
displaying the zero point when the
key is pressed.
When the Application Measurement function is selected, the
key does not wait for stability.
Menu Map item [39]
Print and tare operate without waiting for stabilization.
Menu Map item [40]
Print and tare operate after the stability symbol illuminates.
Note: When the balance is in the
mode and waiting for stability to be determined, -
--- is displayed when the
key is pressed.
Press the
key to disable this function. Taring is not executed after the key is pressed.
Once the
key is pressed, data is not output until the and symbol has illuminated and the
balance is stable. Data is output only after the stability symbol is lit.

Table of Contents