2.5.5 Calibration Check Using External Weights
1) Verify that the balance is in mass display and unload the sample from the pan.
2) Press the
key once to display .
(If
is not displayed, select Menu Map item [4].)
3) Press the
key.
The value of the correct calibration mass to be loaded is displayed and blinks.
Changing the Calibration Mass to be Used
Pressing the key allows changes to the weight value. Modify the value
using the
key and key, then press the key. To
interrupt modification, press the
key.
4) Load the indicated calibration mass and press the key.
The zero display blinks.
5) Unload the weight from the pan and press the
key.
The display changes to the
xxx display. (xxx indicates a numeric value)
6) To change the
value to zero, press the key.
Press the key to avoid changing the value to zero.
Pressing the
key interrupts calibration and does not change this value
to zero.
Note: Changing the
value to zero is equivalent to performing span calibration.