Appendix 3. Error Code Table
(1)Error that can occur in the Weighing Mode
ERROR 01
Reason for Error
Data stored in the internal memory was deleted due to electric shock.
Troubleshoot
Do the necessary designations in the Set Mode.
ERROR Lc
Reason for Error
The load cell is not connected properly or there is a problem in
the A/D Converter.
Troubleshoot
Check the connection between the scale and the unit.
UNPASS
Reason for Error
The initial zero range has exceed the F13 intial zero range in the Set Mode.
Troubleshoot
Check the weight on scale, and change the F13 value range in
the Set Mode.
ERROR 04
Reason for Error
A key was pressed for a long time or there is a problem in the
key part.
Troubleshoot
Do the key test in Test Mode 1, and if this message continues to appear without any
actual problems, contact the A/S department of our company.
OVER
Reason for Error
A key was pressed for a long time or there is a problem in the
key part.
Troubleshoot
Do the key test in Test Mode 1, and if this message continues to appear without any
actual problems, contact the A/S department of our company.