33
4.1.6 Linearity Adjust (Menu Code 8160)
(Calibration MENU -> 1. Calibration -> 6. Linearity Adjust)
You can re-adjust the med-range weight level for precise calibration.
①
For 5kg weight, input “5”key and press “print”
②
Clear the tray and press “PRINT”, then “Wait4~Wait0” will display
③
Clear the tray and press “PRINT”, then “Wait4~Wait0” will display
③
Put 5kg on the tray the press “PRINT” after “Wait4~Wait0”.
③
Put 15kg on the tray the press “PRINT” after “Wait4~Wait0” exit menu.
8100
CAL
ModE
4. Gravity Constant
5. Percent Calibration
6. Linearity Adjust
LoAd
77407
88137
- Place 5.000 kg on the platter.
- Press PRINT when ready.
Mid
25802
36532
- Place 5.000 kg on the platter.
- Press PRINT when ready.
8160
CAL
ModE
Use Weight :[ 5] kg
Full Capa Weight:15.000 kg
ULoAd
10731
10731
- Remove all weight.
- Press PRINT when ready.