37
4.2.5 Hysteresys Calibration (Menu Code 8185)
(Calibration MENU -> 1. Calibration -> 8. Factory Setting -> 5. Hysteresys Calibration)
①
Selecting “Hysteresys Calibration”
②
For 2kg Mid-weight, input “2”key and press “print”
③
Empty tray and press “PRINT”
④
While calibrating zero display shows “Wait4” ~ “Wait0” and follow next message for Mid
Calibration.
⑤
Put on the Weight for Mid-weight. Capacities then press “PRINT”
⑥
While calibrating Mid-weight display shows “Wait4” ~ “Wait0” and follow next message for Span
Calibration.
8185
CAL ModE
USE WEIGHT :[ 2] kg
Full Capa Weight : 6.000 kg
ULoAd
0
27770
- Remove all weight.
- Press PRINT when ready.
WAit4
27770
27770
- Remove all weight.
- Press PRINT when ready.
Mid U
0
27770
- Place 2.000 kg on the platter.
- Press PRINT when ready.
WAit4
28850
55510
- Place 2.000 kg on the platter.
- Press PRINT when ready.
LoAd
28850
55510
- Place 6.000 kg on the platter.
- Press PRINT when ready.