18
4.7 Hold function
4.7.1 In normal weighing mode, press key to average the unstable weight
readings, and get a value to display for 3 seconds.
The AVG. character flickers for 2 seconds and then it displays the weight. (The
weight is an average weight value)
The weight value will be kept for 3 seconds and return to normal weighing mode.
4.7.2 Long press to extend or auxiliary the display. ( Approval Ver. only)
To increase the number or move to next choice.
Auxiliary indicating device (600 g , 6000 g )
The balance with verification scale interval e= 1x10
P
k
P g are equipped with an
auxiliary indicating device having a display view d = 0.1 e.
Balances with auxiliary indicating device and and ct( (metric carat) enabled as a
selectable weighing unit are to be regarded as dual-range instruments with manual
changeover, and they must be marked with Max, Min e=, d= and T = in both g
(gram) and ct (metric carat).
Extended indicating device ( 300 g, 1500 g , 3000 g )
The balances with verification scale interval e≠1x10
P
k
P g are equipped with an
Extended indicating device, which displays the weight with d = 0.1 e for 5 seconds,
when activated.