83
Temperature Calibration (IR Sensors)
Note:
The Intelli-Temp Calibrator Kit (CEM part # 907330 – 120V or 907340 – 240V) which includes the calibrator and
adaptor is required to perform an IR calibration on the instrument. The Intelli-Temp Calibrator emits a temperature of
130 ºC.
1. The instrument must be plugged into an electrical outlet and turned on.
2. Remove the plastic floor mat from the floor of the instrument.
3. Plug the Intelli-Temp calibrator into an electrical outlet close enough to the instrument for it to be placed on the
instrument floor. Permit the calibrator to warm until the “Ready” light is illuminated.
4. With the MARS 6 home screen displayed, select the System Menu icon at the bottom right corner of the screen.
5. Select “Tools.”

Other manuals for CEM MARS 6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM MARS 6 and is the answer not in the manual?

CEM MARS 6 Specifications

General IconGeneral
TypeMicrowave Digestion System
Power Output1600 W
Temperature ControlYes
Pressure ControlYes
Vessel Capacityup to 40 vessels
DisplayTouchscreen
Operating Pressureup to 800 psi
Temperature Rangeup to 300°C
Temperature Accuracy± 1 °C
Power Supply220-240 V, 50/60 Hz
Safety FeaturesOver-temperature and over-pressure protection
Power Consumption1600W
Cooling RateForced air cooling

Related product manuals