4. With the MARS 6 home screen displayed, select the System Menu icon at the bottom right corner of the screen.
5. Select “Tools.” The inner and outer IR temperatures will be displayed on the “Status” menu.
6. Ensure that the “Ready” light on the Intelli-Temp is illuminated. Place the calibrator adaptor over the inner and
outer IR sensors with the calibrator support ring positioned directly over the sensor being calibrated. Place the
Intelli-Temp calibrator over the adaptor (calibrator positioned upsode down with the opening mating with the
opening of the adaptor).
7. Wait for the temperature to stabilize.
8. The temperature should read 130 ºC ±3 ºC. If it is outside this range, perform the Temperature Calibration for IR
sensors.
Pressure Calibration
Pressure calibration should be performed only by a CEM Certified Service Representative. Contact CEM Service
(800-726-5551).