Specifications:
Overall Dimensions 25 in. (63.5 cm) height x 21 in. (53.3 cm) width x 25 in. (63.5 cm) depth
Weight 140 lbs (63.6 kg)
Touch Screen 7 in. (800 x 480) TFT-LED glass capacitive touch screen display
One Touch A combination of vessel recognition and vessel counting sensor technology,
software technology, and applications knowledge that enables a user to
select a One Touch method that matches the sample type. Based on the
sample type, One Touch Technology determines the vessel type and count,
digestion temperature, ramp and hold times, and microwave power input.
PowerMAX Power control technology provides the optimum amount of energy to the
sample to ensure complete digestions
Ports 5 USB (1 side, 4 rear), 2 Ethernet, 1 USB-B, 1 RS-232
Sensors All sensors, including pressure and temperature sensing devices located
within the microwave cavity, are microwave-transparent or shielded to
ensure accurate readings and to eliminate arcing (ignition) hazards.
Languages Software available in English, German, French, Italian, Spanish, Chinese and
Japanese
Sample Stirring In-vessel magnetic stirring of samples
Turntable Design PerfectCircle™ design provides absolute radial symmetry. Turntable operates
in alternating or continuous mode.
Inlet/Outlet Ports Standard 0.500” I.D. port or optional 0.3125” I.D. ports for 0.250” (6mm)
tubing
Microwave Cavity Heavy-duty, multi-layer Teflon® coating
Electrical Requirements 208/230 VAC (207-253 VAC), 60 HZ, 15A @ 208 VAC
220/240 VAC (202-250 VAC) 50 Hz, 15A @ 220 VAC
Magnetron Frequency 2450 MHz
Power Output 1800 watts – continuous power available at all power levels to provide
more control for reactions (IEC 705 Method – 1988)
Magnetron Protection Solid-state isolator (US patent 4,835,354) to protect magnetron from reflected
energy, ensuring constant power output
Speakers 8 , 2 W, 86 dB
Printer Optional on-board thermal printer and USB-B compatible printer port
Safety Features Three independent door safety interlocks, including an interlock monitoring
system plus three independent thermal switches, are used in each instrument
to prevent instrument operation and microwave emissions in case of
improper door closure or misalignment. The instrument complies with HHS
standards under 21 CFR, Part 1030.10, Subparts (C)(1), (C)(2) and (C)(3).
Reactiguard continuously monitors the cavity and disables the magnetron if
disturbances occur inside the cavity.
Emissions and Safety Approvals Conforms to Globally Harmonized EN61010-1 Standard for Safety
Requirements for Electric Equipment for Measurement, Control and
Laboratory Use Part 1: General Requirements (CAN/CSA-C22.2 No.
1010.1-1992).
Patents CEM microwave systems and vessel designs may be covered by any one of
the following US patents: 04835354, 04080168, 05369034, 04672996,
RE034373, 05230865, 04877624, 04672996, 05206479, 05427741. Other
patents pending.
Teflon is a registered trademark of DuPont.
©2012 CEM Corporation

Other manuals for CEM MARS 6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM MARS 6 and is the answer not in the manual?

CEM MARS 6 Specifications

General IconGeneral
TypeMicrowave Digestion System
Power Output1600 W
Temperature ControlYes
Pressure ControlYes
Vessel Capacityup to 40 vessels
DisplayTouchscreen
Operating Pressureup to 800 psi
Temperature Rangeup to 300°C
Temperature Accuracy± 1 °C
Power Supply220-240 V, 50/60 Hz
Safety FeaturesOver-temperature and over-pressure protection
Power Consumption1600W
Cooling RateForced air cooling