EasyManua.ls Logo

Christ ALPHA 2-4 LDplus - Safety Devices; Procedures in the Event of Hazards and Accidents

Christ ALPHA 2-4 LDplus
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Freeze-dryer Alpha 1-4 LDplus
Freeze-dryer Alpha 2-4 LDplus
3 Safety
Version 11/2006, Rev. 2.2 of 17/10/2016 sb
27 / 102
Translation of the original operating manual
Pos: 66 /200 Chri st/360 GT-BA Labor-Pilot (ST ANDARDMODU LE)/030 Sic herheit/030-001 0 Sicherheit La bor/030-007 0 Sicherheitseinri chtungen------------------------------------------------------ @ 25\mod_14 04983309110_ 68.docx @ 1 83369 @ 2 @ 1
3.7 Safety devices
Pos: 67 /200 Chri st/360 GT-BA Labor-Pilot (ST ANDARDMODU LE)/030 Sic herheit/030-001 0 Sicherheit La bor/030-007 0-0010 Syste mkontrolle LDplus @ 31\mod_14085155 08856_68.doc x @ 209772 @ 3 @ 1
3.7.1 System check
An internal system check system monitors the data transfer and sensor
signals with regard to plausibility. Errors are detected by continuous self-
monitoring of the system. Error messages are displayed in the main
window under "Process & equipment messages" ((see chapter 7.2 -
"Process and equipment messages")).
Pos: 68 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 69 /200 Chri st/360 GT-BA Labor-Pil ot (STANDARDM ODULE)/030 Sicherheit/03 0-0010 Sicher heit Labor/030- 0070-0020 Sch utzleiterprüfung Al pha - Ga mma, Epsilon 1- 4_2-4 @ 25\ mod_14049833 11463_68.doc x @ 183397 @ 3 @ 1
3.7.2 Earth conductor check
For the earth conductor check, there is an equipotential bonding screw on
the rear panel of the freeze-dryer. An earth conductor check can be carried
out with the aid of a suitable measuring instrument.
Pos: 70 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 71 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 72 /200 Chri st/360 GT-BA Labor-Pilot (ST ANDARDMODU LE)/030 Sic herheit/030-001 0 Sicherheit La bor/030-008 0 Verhalten bei G efahren und U nfällen Alpha- Gamma, Epsilo n 1-4_2-4 @ 25\mod_1404 983312451_68.doc x @ 183411 @ 2 @ 1
3.8 Procedures in the event of hazards and accidents
Hazardous electrical incident:
Set the control switch to the "0" position in order to interrupt the power
supply completely.
Fire:
A fire in the electrical control system must be extinguished with a CO
2
fire extinguisher!
Burning oil must be extinguished with a CO
2
fire extinguisher or powder
fire extinguisher!
Electric shock:
While ensuring your own safety, interrupt the circuit as quickly as
possible (control switch). Keep the affected persons warm and calm.
Get medical attention immediately! Check consciousness and
breathing continuously. In the case of unconsciousness of lack of
normal breathing, perform cardiopulmonary resuscitation (CPR).
Burns:
Cool small-area burns (e.g. finger) immediately with cold water for
approximately 2 minutes.
Do not cool if larger areas of the body surface are burnt since there is a
risk of hypothermia.
Cover the burns loosely and in a sterile manner (e.g. with sterile
dressing).
Keep the affected persons warm and calm.
IF IN DOUBT, CALL THE EMERGENCY PHYSICIAN (AMBULANCE)!
Pos: 73 /010 Uni versalmodule/S eitenwechsel @ 0\mod_120 2116244312_ 0.docx @ 105 @ @ 1

Table of Contents