7 Malfunctions and error correction
Version 11/2006, Rev. 2.2 of 17/10/2016 • sb
Translation of the original operating manual
Pos: 143 /200 Chris t/360 GT-B A Labor-Pilot ( STANDARDMOD ULE)/070 St örungen und F ehlersuche/070 Störungen un d Fehlersuche=== ======= ======= ========== @ 25\m od_14049840 56039_68.doc x @ 184111 @ 1 @ 1
7 Malfunctions and error correction
Malfunctions are displayed in the dialog box "Process & equipment
messages" (see chapter 6.5.3.3 - "Process and equipment information").
An acoustic signal sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapter).
• Acknowledge the error message.
Pos: 144 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 145 /200 Chris t/360 GT-B A Labor-Pilot ( STANDARDMOD ULE)/070 St örungen und F ehlersuche/070- 0010 Allge meine Störungen LD plus @ 31\m od_14084382 56415_68.doc x @ 209704 @ 2 @ 1
7.1 General malfunctions
Pos: 146 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 147 /200 Chris t/360 GT-B A Labor-Pilot ( STANDARDMOD ULE)/070 St örungen und F ehlersuche/070- 0010-0010 Ne tzausfall @ 25\mod_140498 4058964_68. docx @ 184153 @ 3 @ 1
7.1.1 Power failure
The control system continues with the process after a power failure. The
preselected conditions remain saved even during a process run.
In the event of a power failure in the drying phase, the batch may become
unusable. Whether the batch can be saved or not depends on the drying
phase in which the product was when the power failure occurred.
• In the final drying phase, the product has reached a residual moisture
content of approx. 5%. Below this value, the product is generally not
damaged even if the power failure lasts for a longer period of time.
• If the product is in the main drying phase, we recommend aerating the
unit, removing the product, and storing it in a deep-freeze. The
defrosted condensate must be drained off prior to the next start.
Pos: 148 /010 Uni versalmodule/ Leerz eile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 149 /200 Chris t/360 GT-B A Labor-Pilot ( STANDARDMOD ULE)/070 St örungen und F ehlersuche/070- 0010-0020 K ein ausreichend es Vakuum @ 2 5\mod_140498 4060039_68.d ocx @ 1841 67 @ 3 @ 1
7.1.2 Insufficient vacuum
The vacuum checks must be carried out when the ice condenser is frozen.
Pos: 150 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 151 /200 Chris t/360 GT-B A Labor-Pilot ( STANDARDMOD ULE)/070 St örungen und F ehlersuche/070- 0010-0020-0 010 Kleinflansc hverbindungen @ 25\mod_ 1404984061086_68. docx @ 184 181 @ 4 @ 1
7.1.2.1 Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
• Seal the connection with the clamping ring by tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.