Freeze-dryer Alpha 1-4 LDplus
Freeze-dryer Alpha 2-4 LDplus
7 Malfunctions and error correction
Version 11/2006, Rev. 2.2 of 17/10/2016 sb
57 / 102
Translation of the original operating manual
Error message
Possible cause
Remedy
Overpressure of refrigeration unit 2
The overpressure switch of the
refrigeration unit has tripped.
Let the freeze-dryer cool down.
Ensure sufficient air circulation
(see chapter 7.1.3 - "Insufficient
ice condenser temperature").
Ice condenser overtemperature
Ice condenser temperature > +65°C. All
of the units will be switched off and the
freeze-dryer will be automatically set to
standby.
The defrosting process has
been performed with the cover
closed.
Remove the cover and let the
freeze-dryer cool down.
Ice condenser temperature > 20°C
The ice condenser temperature rises to
a non-permissible value during the
freeze-drying process. The pressure
control valve will close automatically in
order to prevent the vacuum pump from
being damaged.
Insufficient cooling.
Observe the behaviour of the
refrigeration system. If
necessary, inform the service
department (see chapter 7.3 -
"Service contact").
Defective ice condenser temperature
sensor
The temperature sensor is not
connected or it is defective.
Inform the service department
(see chapter 7.3 - "Service
contact").
Defective temperature sensor of the
defrosting heater
The temperature sensor is not
connected or it is defective.
Inform the service department
(see chapter 7.3 - "Service
contact").
Mains power failure (see chapter 7.1.1
- "Power failure")
The power supply has been
interrupted.
The mains power connector is
not connected.
The fuses have tripped.
The mains power switch has
been switched off.
Check the mains fuse.
Connect the mains power
connector firmly.
Check the customer-provided
fuses.
Switch the mains power switch
on.
Default settings loaded
All of the settings have been reset to the
delivery state of the freeze-dryer.
Error in the parameter memory.
Manual execution in the service
menu.
No measure required.
IO communication error (0x20)
Error in the LDplus IO module.
Inform the service department
(see chapter 7.3 - "Service
contact").