Description of Panel
4-5
KEY LOCK:
It locks the key. The default setting is OFF. Switch to MEASURE DISPLAY or BIN
COUNT or COMPARE COUNT the word <LOCK> will appear at upper right of test screen
when it is on. To disable it, press [F1], [F4], and [SYSTEM SETUP].
CONTRAST:
LCD contrast adjustment range is 0 13. The default setting is 7”.
SOUND MODE:
FAIL: It beeps when detected no good product during performing the measurement of
BIN or COMPARE.
PASS: It beeps when detected good product during performing the measurement of BIN
or COMPARE.
The default setting is FAIL.
ALARM MODE:
PULSE: It sets the warning to a short beep during good/no good product judgment.
CONTINUE: It sets the warning to a continuous long beep during good/no good product
judgment.
The default setting is PULSE.
TRIG. DELAY:
It is to adjust the time delayed for measurment when the meter receives the trigger signal.
The range is from 0 to 1000mS, and the default setting is 0000mS. All trigger modes will
be affected by this setting.
TRIG. EDGE:
It sets the RISING and FALLING edge trigger. The default setting is FALLING edge
trigger.
HANDLER MODE:
CLEAR: When the Handler interface is in use, it will clear the output signal (PASS or
FAIL) of previous tested result before measuring.
HOLD: When the Handler interface is in use, the output signal (PASS or FAIL) of
tested result will remain until the next test result shows otherwise.
The default setting is CLEAR.
MEAS. DELAY:
The time delayed before measuring each time, the range is 0.000 100 seconds. The
setting method is by using [F1] key (DIGIT UP) or [F2] key (DIGIT DOWN) to adjust.
Its interval time in 5mS~100mS is 5mS, 0.1~100S is 0.1S. Please see section 4.5
Reference Data for Operation.
LINE FREQ.:
It sets the line frequency to 50Hz or 60Hz for AC110V/220V power source. The default
setting is 60Hz.
GPIB ADDRESS:
It sets the GPIB interface address. The default setting is 17 and the range is from 01 to
30.
BAUDRATE:
It sets baud rate of RS232 serial port. There are 1200, 2400, 4800, 9600, 19200, 38400
six selections. By using [F4] key is to switch setting value selection of up/down page.