Milliohm Meter 16502 Users Manual
5-8
Return: The query returns DEV or PCNT.
(6) Command: CALCulate:BINNing: BIN(1~8): UPPer <NRf+> {Suffix Unit}
Function: It sets or queries the high limit of BIN.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
DEV parameter range: 0.0000 MOHM(MIN) 200.0000 MAOHM(MAX).
PERCENT parameter range: 0.00(MIN) 999.99(MAX).
Unit: The units of parameter are {MOHM | OHM | KOHM | MAOHM} when
CALCulate:BINNing:MATH:NAME is set to DEV. If there is no giving unit, OHM
will be defined automatically. Interval between paramter and unit is by blank.
Ex.: CALCulate:BINNing:BIN1:UPPer 101.000 KOHM
If CALCulate:BINNing:MATH:NAME setting is PCNT, there is no parameter unit.
Return: The query returns <NR2> suffix unit. Ex.: 101.000 KOHM or 100.00 %.
(7) Command: CALCulate:BINNing:BIN(1~8):LOWer <NRf+> {Suffix Unit}
Function: It sets or queries the low limit of BIN.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
DEV parameter range: 0.0000 MOHM(MIN) 200.0000 MAOHM(MAX)
PERCENT parameter range: 0.00(MIN) 999.99(MAX).
Unit: The units of parameter are {MOHM | OHM | KOHM | MAOHM} when
CALCulate:BINNing:MATH:NAME is set to DEV. If there is no giving unit, OHM
will be defined automatically. Interval between paramter and unit is by blank.
Ex.: CALCulate:BINNing:BIN1:LOWer 99.000 KOHM
If CALCulate:BINNing:MATH:NAME setting is PCNT, there is no parameter unit.
Return: The query returns <NR2> suffix unit. Ex.: 99.000 KOHM or 100.00 %.
(8) Command: CALCulate:BINNing:NOMInal{<numeric_value|MIN|MAX} [Suffix Unit]
Function: It sets or queries nominal of BIN SORT.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
Parameter range: 0.0000 MOHM(MIN) 200.0000 MAOHM(MAX)
Unit: It defines unit of setting parameter {MOHM | OHM | KOHM | MAOHM}, if no
the command of unit, OHM will be defined automatically. Interval between
paramter and units is by blank.
Ex.: CALCulate:BINNing:NOMInal 100.000 MOHM
Return: The query returns <NR2> with suffix unit.
Ex.: 100.0000 KOHM
(9) Command: CALCulate:BINNing:STATe {OFF | ON | 0 | 1 }
Function: It sets or queries if BINNING function is enabled.
Parameter: OFF (0) Disable BINNING function
ON (1) Enable BINNING function
Return: The query returns 0 or 1.
(10) Command: CALCulate:COMPare:CLEAr
Function: It clears STBY state of comparison result.
Parameter: None
Return: None
(11) Command: CALCulate:COMPare:RESUlt?
Function: It returns comparison result.
Parameter: None
Return: The query returns <NR1> +0 ~ +11,
+0 COMPARE OFF or comparison result is LO.
+9 Comparsion result is HI.
+10 Comparsion result is PASS.
+11 STBY state
(12) Command: CALCulate:COMPare:MATH:EXPRession:CATalog?