Descriptions of GPIB Interface (Commands Same as RS232 Interface)
5-9
Function: It returns numeric value after mathematics by setting (CALCulate:
COMPare:MATH: EXPRession:NAME)
Parameter: None
Return: The query returns <NR3>.
(13) Command: CALCulate:COMPare:MATH: EXPRession: NAME { DEV|PCNT}
Function: It sets or queries the expression of numeric value.
Parameter: DEV The expression of numeric value is absolute value.
PCNT The expression of numeric value is percentage value.
Return: The query returns DEV or PCNT.
(14) Command: CALCulate:COMPare:MATH:STATe {OFF | ON | 0 | 1 }
Function: It sets or queries CALCulate:COMPare:MATH:EXPRession:NAME
definition in mathematics process.
Parameter: ON (1) Enable mathematics process.
OFF(0) Disable mathematics process.
Return: The query returns 0 or 1.
(15) Command: CALCulate:COMPare:LIMit:STATe {OFF | ON | 0 | 1}
Function: It sets or queries comparison function is enabled.
Parameter: OFF (0) Disable comparison function.
ON (1) Enable comparison function.
Return: The query returns <NR1> 0 or 1.
(16) Command: CALCulate:COMPare:LIMit:NOMInal <NRf+> {Suffix Unit}
Function: It sets or queries nominal of comparison function.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
Parameter range: 0.0000 MOHM(MIN) ~ 200.0000 MAOHM(MAX)
Unit: It defines unit of parameter {MOHM | OHM | KOHM | MAOHM}, if no
the command of unit, OHM will be defined automatically. Interval between
paramter and units is by blank.
Ex.: CALCulate:COMPare:LIMit:NOMInal 100.000 KOHM
Return: The query returns <NR2> with suffix unit.
Ex.: 100.0000 KOHM
(17) Command: CALCulate:COMPare:LIMit:LOWer <NRf+> {Suffix Unit}
Function: It sets or queries the low limit of comparison function.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
DEV parameter range: 0.0000 MOHM(MIN) ~ 200.0000 MAOHM(MAX)
PERCENT parameter range: 0.00(MIN) ~ 999.99(MAX)
Unit: The units of parameter are {MOHM | OHM | KOHM | MAOHM} when
CALCulate:COMPare:MATH:EXPression:NAME is set to DEV. If there is no
giving unit, OHM will be defined automatically. Interval between paramter and
unit is by blank.
Ex.: CALCulate:COMPare:LIMit: LOWer 99.000 KOHM
If CALCulate:COMPare:MATH:EXPression:NAME setting is PCNT, there is
no parameter unit.
Return: The query returns <NR2> with suffix unit.
Ex.: 99. 000 KOHM or 100.00 %.
(18) Command: CALCulate:COMPare:LIMit:UPPer <NRf+> {Suffix Unit }
Function: It sets or queries the high limit of comparison function.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
DEV parameter range: 0.0000 MOHM(MIN) ~ 200.0000
MAOHM(MAX).
PERCENT parameter range: 0.00(MIN) ~ 999.99(MAX).
Unit: The units of parameter are {MOHM | OHM | KOHM | MAOHM} when
CALCulate:COMPare:MATH:EXPression:NAME is set to DEV. If there is no