Milliohm Meter 16502 Users Manual
5-10
giving unit, OHM will be defined automatically. Interval between paramter and
unit is by blank.
Ex.: CALCulate:COMPare:LIMit: UPPer 101.000 KOHM
If CALCulate:COMPare:MATH:EXPression:NAME setting is PCNT, there is
no parameter unit.
Return: The query returns <NR2> with suffix unit.
Ex.: 101.0000 KOHM
TEMPerature Command Set
(1) Command: TEMPerature: UNIT {DEGC|DEGF}
Function: It sets or queries the unit of temperature value.
Parameter: DEGC : °C, DEGF: °F
Return: The query returns DEGC or DEGF.
(2) Command: TEMPerature:ATEMP:MODE {OFF|AUTO|MAN}
Function: It sets ambient temperature mode of conversion and correction functions.
Parameter: OFF means temperature correction doesnt be executed under correction
function as well as the current ambient temperature equals to the initial
temp. setting under conversion function.
AUTO means measuring the current ambient temperature by 16502
optional tester. (If there is no optional device, thus AUTO mode is
disabled.)
MAN means inputting the current ambient temperature by users.
Return: The query returns OFF, AUTO or MAN.
(3) Command: TEMPerature:ATEMP:INITial <NRf+>
Function: It sets or queries the initial temperature value of conversion function.
Parameter: <NR2> | MINimum | MAXimum parameter is acceptable.
TEMPerature: When UNIT is DEGC, the value setting range -10(MIN) ~
99.9(MAX).
TEMPerature: When UNIT is DEGF, the value setting range 14(MIN) ~
211.8(MAX).
Return: The query returns <NR2> Ex.: +25.0.
(4) Command: TEMPerature:ATEMP[:CURRent] <NRf+>
Function: It sets or queries the current ambient temperature.
Parameter: When TEMPerature:ATEMP:MODE is MAN for setting the current ambient
temperature value. <NR2> | MINimum | MAXimum parameter is
acceptable.
TEMPerature: When UNIT is DEGC, the value setting range -10(MIN) ~
99.9(MAX).
TEMPerature: When UNIT is DEGF, the value setting range 14(MIN) ~
211.8(MAX).
Return: The query returns <NR2>. Ex.: +25.0.
(5) Command: TEMPerature:RESistance[:INITial] <NRf+> {Suffix Unit}
Function: It sets or queries DUT’s resistance under initial temperature of conversion
function.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
Parameter range: 0.0000 MOHM(MIN) ~ 999.999 MAOHM(MAX)
Unit: It defines the unit of setting parameter {MOHM | OHM | KOHM | MAOHM}, if
there is no command of unit, OHM will be defined automatically. Interval between
paramter and units is by blank.
Ex.: TEMPerature:RESistance:INITial 101.000 KOHM
Return: The query returns <NR2> with suffix unit. Ex.: 101.0000 KOHM.
(6) Command: TEMPerature:CONStant <NRf+>