Descriptions of GPIB Interface (Commands Same as RS232 Interface)
5-11
Function: It sets or queries the coefficient of temperature conversion function.
Parameter: <NR2> | MINimum | MAXimum parameter is acceptable.
Range: 0.0(MIN) ~99.9(MAX)
Return: The query returns <NR2>.
(7) Command: TEMPerature:CORRect <NRf+>
Function: It sets or queries the reference temperature of correction function.
Parameter: <NR2> | MINimum | MAXimum parameter is acceptable.
TEMPerature: When UNIT is DEGC, the value setting range -10(MIN) ~
99.9(MAX).
TEMPerature: When UNIT is DEGF, the value setting range 14(MIN) ~
211.8(MAX).
Return: The query returns <NR2>. Ex.: +25.0.
(8) Command: TEMPerature:TCOEF <NRf+>
Function: It sets or queries thermal coefficient parameter of correction function.
Parameter: <NR1> | MINimum | MAXimum parameter is acceptable.
Range 1(MIN) ~9999(MAX)
Return: The query returns <NR1>.
(9) Command: TEMPerature:CONVersion:MODE { ABS|DEV|0|1|}
Function: It sets the temperature display mode of conversion function.
Parameter: ABS or 0 The temperature of conversion displays T.
DEV or 1 The temperature of conversion displays T.
Return: The query returns ABS or DEV.
(10) Command: TEMPerature:CONVersion[:RESUlt]?
Function: It queries the temperature value of conversion.
Parameter: None
Return: The query returns <NR3>.
(11) Command: TEMPerature:PROBE {0|1|}
Function: It sets temperature probe type.
Parameter: 0 PT100
1 PT500
RSCAN Command Set
(1) Command: RSCAN:LIMit:NOMInal <NRf+> {Suffix Unit}
Function: It sets or queries nominal value of RSCAN function STEP and BL
parameter.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
Parameter range: 0.0000 MOHM(MIN) ~ 200.0000 MAOHM(MAX).
Unit: It defines the unit of setting parameter {MOHM | OHM | KOHM | MAOHM}, if
there is no giving unit, OHM will be defined automatically. Interval between
paramter and unit is by blank.
Ex.: RSCAN:LIMit:NOMInal 100.000 KOHM
Return: The query returns <NR2> with suffix unit. Ex.: 100.0000 KOHM.
(2) Command: RSCAN:LIMit:STEP:MATH:NAME {DEV |PCNT}
Function: It sets or queries upper/lower limit setting value expression for STEP1~3.
Parameter: DEV indicates the expression is absolute value.
PCNT indicates the expression is percentage value.
Return: The query returns DEV or PCNT.
(3) Command: RSCAN:LIMit:STEP{1|2|3}:UPPer <NRf+> {Suffix Unit}
Function: It sets or queries the upper limit for STEP1~3.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
DEV parameter range: 0.0000 MOHM(MIN) ~ 200.0000 MAOHM(MAX).