Milliohm Meter 16502 Users Manual
5-12
PERCENT parameter range: 0.00(MIN) ~ 999.99(MAX).
Unit: The units of parameter can be set to {MOHM | OHM | KOHM | MAOHM} when
RSCAN:LIMit:STEP:MATH:NAME is set to DEV. If there is no giving unit, OHM
will be defined automatically. Interval between paramter and unit is by blank.
Ex.: RSCAN:LIMit:STEP1:UPPer 101.000 KOHM
When RSCAN:LIMit:STEP:MATH:NAME setting is PCNT, there is no parameter
unit.
Return: The query returns <NR2> with suffix unit. Ex.: 101.000 KOHM or 100.00 %.
(4) Command: RSCAN:LIMit:STEP{1|2|3}:LOWer <NRf+> {Suffix Unit}
Function: It sets or queries the lower limit for STEP1~3.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
DEV parameter range: 0.0000 MOHM(MIN) ~ 200.0000 MAOHM(MAX)
PERCENT parameter range: 0.00(MIN) ~ 999.99(MAX)
Unit: The units of parameter can set to {MOHM | OHM | KOHM | MAOHM} when
RSCAN:LIMit:STEP:MATH:NAME is set to DEV. If there is no giving unit, OHM
will be defined automatically. Interval between paramter and unit is by blank.
Ex.: RSCAN:LIMit:STEP1:LOWer 100.000 KOHM
If RSCAN:LIMit:STEP:MATH:NAME setting is PCNT, there is no parameter unit.
Return: The query returns <NR2> with suffix unit. EX.: 100.000 KOHM or 100.00 %.
(5) Command: RSCAN:LIMit:BL:MATH:NAME {DEV |PCNT}
Function: It sets or queries BL upper/lower limit setting expression.
Parameter: DEV indicates the expression is absolute value.
PCNT indicates the expression is percentage value.
Return: The query returns DEV or PCNT.
(6) Command: RSCAN:LIMit:BL:UPPer <NRf+> {Suffix Unit}
Function: It sets or queries BL upper limit.
Parameter: <NRf> | MINimum | MAXimum parameter is acceptable.
DEV parameter range: 0.0000 MOHM(MIN) ~ 200.0000 MAOHM(MAX)
PERCENT parameter range: 0.00(MIN) ~ 999.99(MAX)
Unit: The units of parameter can set to {MOHM | OHM | KOHM | MAOHM} when
RSCAN:LIMit:BL:MATH:NAME is set to DEV. If there is no giving unit, OHM will
be defined automatically. Interval between paramter and unit is by blank.
Ex.: RSCAN:LIMit:BL:UPPer 100.000 KOHM
If RSCAN:LIMit:BL:MATH:NAME setting is PCNT, there is no parameter unit.
Return: The query returns <NR2> with suffix unit. Ex.: 100.000 KOHM or 100.00 %.
(7) Command: RSCAN:TEST:STEP{1|2|3}:PIN <numeric_value>
Function: It sets or queries test pin for STEP1~3.
Parameter: numeric_value 0~6
0: 0-0(indicates STEP no testing)
1:1-4
2:2-4
3:3-4
4:1-2
5:2-3
6:1-3
Return: numeric_value0~6
(8) Command: RSCAN:MEASure:STEP{1|2|3}?
Function: It queries measurement resistance for STEP1~3.
Parameter: None
Return: The query returns <NR3>.
(9) Command: RSCAN:MEASure:STEP{1|2|3}:RESUlt?
Function: It queries measurement result for STEP1~3.