Descriptions of GPIB Interface (Commands Same as RS232 Interface)
5-13
Parameter: None
Return: The query returns <NR1>. 0, 9, 10, 11.
0 measurement result is LO.
9 measurement result is HI.
10 measurement result is GO.
11 is in STBY status not process measurement.
(10) Command: RSCAN:MEASure:BL?
Function: It queries BL value.
Parameter: None
Return: The query returns <NR3>.
(11) Command: RSCAN:MEASure:BL:RESUlt?
Function: It queries BL measurement result.
Parameter: None
Return: The query returns <NR1>, 9, 10, 11.
9 measurement result is HI.
10 measurement result is GO.
11 is in STBY status not process measurement.
(12) Command: RSCAN:MEASure:RESUlt?
Function: It queries measurement result.
Parameter: None
Return: The query returns <NR1> 9, 10, 11.
9 measurement result is FAIL.
10 measurement result is PASS.
11 is in STBY status not process measurement.
SENSe Command Set
(1) Command: SENSe:AVERage:COUNt < NR1>
Function: It sets or queries average of measuring.
Parameter: <NR1> 1 ~ 10
Return: <NR1> 1 ~ 10
(2) Command: SENSe:ZERO:STATe { OFF | ON | 0 | 1}
Function: It sets or queries calculation function of ZERO.
Parameter: OFF(0) Disable calculation function of ZERO.
ON(1) Enable calculation function of ZERO.
Return: The query returns 0 or 1.
(3) Command: SENSe:ZERO:DATA?
Function: It queries calculation function of ZERO.
Parameter: None
Return: The query returns <NR3>.
(4) Command: SENSe:RANG<NRf+>
Function: It sets or queries the measurement range. When Dry Circuit Test is enabled,
the range is not in allowed range. The range isn’t change and Error
displayed.
Parameter: <NR1> | MINimum | MAXimum parameter is acceptable. Parameter range
0(20mΩ)|1(200mΩ)|2(2Ω)|3(20Ω)|4(200Ω)|5(2KΩ)|6(20KΩ|7(200KΩ)|
8(2MΩ)
Return: The query returns <NR1> 0~8.
(5) Command: SENSe:RANG:AUTO {OFF | ON | 0 | 1}
Function: It sets or queries if auto range is enabled.
Parameter: OFF (0) Select the range manually.
ON (1) Select the range automatically.