Milliohm Meter 16502 Users Manual
5-14
Return: The query returns <NR1> 0 or 1.
(6) Command: SENSe:SPEEd { MAX|VFAST|FAST|MEDIum | SLOW}
Function: It sets or queries the measurement speed.
Parameter: MAX The measurement speed is maximum speed.
VFAST The measurement speed is very fast.
FAST The measurement speed is fast.
MEDIUM The measurement speed is medium.
SLOW The measurement speed is slow.
Return: The query returns in the format of string is as MAX, V.FAST, FAST, MEDI or
SLOW.
SOURce Command Set
(1) Command: SOURce:DRY {OFF | ON | 0 | 1}
Function: It sets or queries Dry Circuit Test is enabled.
Parameter: OFF (0) Disable the Dry Circuit Test mode.
ON (1) Enable the Dry Circuit Test mode.
Return: The query returns <NR1> 0 or 1.
(2) Command: SOURce:DRIVe { 0(PULSE+/-) | 1(PULSE +) | 2(PULSE-) | 3(DC +) | 4(DC
-) | 5(DC+ NA) | 6(STBY)}
Function: It sets or queries the mode of DRIVE.
Parameter: 0 is the PULSE +/- mode.
1 is the PULSE + mode.
2 is the PULSE- mode.
3 is the DC+ mode.
4 is the DC- mode.
5 is the DC+ NA mode.
6 is the Stand By mode.
Return: The query returns <NR1> 0~6.
READ Command Set
(1) Command: READ?
Function: It queries the measurement result at present. If the command of
measurement (TRIGger or *TRG) is not executed, a return invalid value will
be produced at the same time as well as an error message (“Data stale”)
also is produced.
Parameter: None
Return: The query returns <NR3>.
TRIGger Command Set
(1) Command: TRIGger:SOURce {INTernal|MANual|EXTernal|BUS|SMT}
Return: {BUS|EXTernal|INTernal|MANual}
Function: It sets or queries the trigger mode at present.
Parameter: INTernal Internal continuous trigger
MANual Manual trigger
EXTernal External trigger mode.
BUS BUS trigger mode.
SMT SMART trigger
Parameter: The query returns INT|MAN |EXT|BUS|SMT
(2) Command: TRIGger:DELay <NR1>
Function: It sets or queries the time delayed for trigger, the range is 0 999ms.
Parameter: Acceptable <NR1>
Return: <NR1>