HIPOT Tester 19051/19052/19053/19054 Quick Start Guide
4
Indication, Alarm
GO: (Short Sound)
NG: W-Arc, W-Hi, W-Lo, IR-Lo, IR-Hi, GFI,
Continuity-fail (Long Sound)
Remote
Connector
Rear Panel 9 Pin
D-type
Connector
Input: Start, Stop, Interrupt (at 11 pin terminal block)
Output: Under test, Pass, Fail
TEST/RESET
Control
Low - active control, (24V open voltage typical).
Input requirements
Input time duration: 20msec. approx.
The above input circuits are not isolated from other
internal circuits.
Options
Interface Card
GP-IB Interface Talk, Listen all function
RS232 (standard
option)
Baud rate: 300 ~ 19200, data bits: 8, stop bit: 1
Ambient Temperature and Relative Humidity
Specifications
range
18 to 28°C (64 to 82°F), ≤ 70% RH.
Operable range Maximum relative humidity 80% for temperature up to
31°C (88°F). Decreasing linearly to 50% relative
humidity at 40° C (104°F)
Altitude up to 2000m.
Indoor use only.
Pollution degree 2
Storage range
-10 to 60°C (-14 to 140°F), ≤ 80% RH.
Installation
Category
CAT II
Power Requirement
Line Voltage AC 100V, 120V, 220V ± 10%, 240V +5 -10%
Frequency 50 or 60 Hz
Power No load: < 100W
Consumption With rated load: 500W max.
Dimension 320W x 105H x 400D mm
Weight 19051, 19052: 14kg approx.
19053, 19054: 15kg approx.
Safety
Ground Bond
Less than 100mΩ at 25Amp, 10sec
Hipot Less than 10mA at Wac 1.5kV, 3sec
Insulation
Resistance
Over 20MΩ at 500V 10sec
Line leakage
current
Less than 3.5mA at 127V, 3sec, normal, reverse