HIPOT Tester 19051/19052/19053/19054 Quick Start Guide
26
JUDG. means during ramp time will
judge high limit under DC mode.
When set this item to OFF, it
means during ramp time won’t
judge high limit under DC mode.
Part No. Not over 13
characters
None Set the product Part No.
Lot No. Not over 13
characters
None Set the product Lot No.
Serial
No.
Not over 13
characters
None Set the product serial no. format,
* means changeable character.
3.7 PROGRAM Setting
3.7.1 Test Procedure Setting
1. Under power on menu, press Function Key PROGRAM and then enter
PROGRAM setting menu as the following:
LOW : 0.001mA
STEP 1 DC
ARC : OFF
UP
VOLT: 0.050kV RAMP : 999.0s
HIGH: 0.500mA FALL : OFF
MORE..
TIME: 3.0s CHK : OFF
ENTER
DWLL: OFF 1 2 3 4 5 6 7 8
SCAN : X X X X X X X X
EXIT
PROCESS STEP RMT LOCK OFST ERR
2. After entering PROGRAM setting menu, use Function Keys UP select
the test procedure want to set, the range is 199.
3. Press ENTER key move the highlighted cursor to the parameter item,
which want to set.
4. Press Function Key MORE.. can switch to other setting menu as the
following.