HIPOT Tester 19051/19052/19053/19054 Quick Start Guide
25
AC-V
Freq.
50/60 60 Set AC-V FREQ. of HIPOT tester
by inputting frequency of AC
source.
GR
CONT.
OFF/KEY/TIME
(0.2sec~99.9sec)
OFF Set grounding continue test no
good function operation mode.
1. When set to OFF, it doesn’t
proceed grounding continue
test.
2. When set to KEY, press
START KEY to proceed
grounding continue test.
3. When set to TIME, GR
CONT operation modes are
as below descriptions.
(1) When users press
START KEY, the
program judge if DUT
connected well by GR
CONT. ON or OFF.
(2) If CR CONT. judge DUT
is connected well then
proceed test
automatically when set
TIME is up.
(3) After the test is ended,
re-judge if CR CONT.
is continue for the
condition of proceeding
test.
Soft.
AGC
ON/OFF ON Set software automatic gain
compensation function is open or
not.
Auto
Range
ON/OFF OFF Set withstand voltage auto-range
function is open or not.
GFI ON/OFF ON Set ground fail interrupt function
AFTER
FAIL
STOP/CONTINUE/R
ESTART
STOP After setting FAIL, it indicates if
stop the test or continue to the
next step or restart.
SCREEN ON/OFF ON Set if show test screen.
SMART
KEY
ON/OFF OFF Set if open parameter memory
function.
RAMP ON/OFF ON When set this item to ON, it