HIPOT Tester 19051/19052/19053/19054 Quick Start Guide
14
it for a long time, please use original material packing and then store it.
For correct test and safety, please keep it from direct sunlight or high
temperature, vibration, humidity and dusty place.
21. Warm up
All functions of the tester are activated when the power switch is turned
on. However, to attain the precision in the specification, please warm
the instrument over 15 minutes.
22. Warning signal of testing
"DANGER – HIGH VOLTAGE TEST IN PROGRESS,
UNAUTHORIZED PERSON KEEP AWAY"
23. Keep test cable away from the panel
Please keep the high voltage cable or the DUT away from the panel at
least 30 cm during operation to avoid the display interference caused
by high-voltage discharge.
24. Notices for connecting automated device
y The grounding system of the device and the automated station
should be connected together.
y Add anti-interference iron core to the high voltage cable and the 2
ends (device output and DUT) of RTN/LOW test cable with
winding at least 1 circle.
y The high voltage and RTN/LOW test cable must be separate from
the control cable.
y The high voltage and RTN/LOW test cable must keep proper
distance from the tester panel.