HIPOT Tester 19051/19052/19053/19054 Quick Start Guide
18
lower than 0.3A. The action time: From judging FAIL to
STOP is stopped.
OUTPUT Switch: When toggles this switch to power symbol,
UNDER TEST output terminal will be short
circuited under test status. When toggles this
switch to voltage symbol, UNDER TEST terminal
outputs 24V under test status. This function can
be used with 3002B or 3002D and is for controlling
valve.
2. VOLTAGE SELECTOR Input Power Supply Range Switch
Changing the tester inputted AC power. Using AC power has four
kinds as below.
a. 90 110V AC
b. 108 132V AC
c. 198 242V AC
d. 216 250V AC
Switching this power switch by applying AC power and notice the
change of fuse.
3. AC LINE: AC power socket and fuse holder.
A tri-cord power and fuse holder. Input AC power, which the tester is
needed from AC power socket. The detailed specification of using
fuse please refers "Chapter 3 - Notice Items Before Using" or
descriptions of rear panel in this manual.
4. GROUND: Safety GND terminal. Please use adaptable implement to
connect this grounding terminal actually. If there is no grounding
actually, the circuit with GND terminal or other instruments connecting
cable with GND terminal is short circuit. The cover of tester may exist
high voltage. This is very dangerous, anyone touch the tester under
the above status may cause damage. Therefore, it is necessary to
connect safety GND terminal to ground.
5. GPIB INTERFACE (Option)
This socket is for optional GIPB interface (IEEE-488-1978).
6. OPTION: This socket is the option PRINTER interface for the tester.
7. FAN: The temperature control fan.
When the temperature reaches 50°C, fan opens automatically. When
the temperature is lower than 45°C, fan stops automatically.