Using Remote Control
5-27
MEASure:VOLTage:HARMonic:ARRay?<space><CRD>,<NR1><P
MT>
Query parameters: <CRD>, VALUE | PERCENT
<NR1>, 1 ~ 4: Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:VOLTAGE:HARMONIC:ARRAY<space><NR2>,<NR2>,
<NR2>,...up to 101 <RMT>
Header off: <NR2>,<NR2>,<NR2>,...up to 101 <RMT>
Separator 0: <NR2>,<NR2>,<NR2>,...up to 101 <RMT>
Separator 1: <NR2>;<NR2>;<NR2>;...up to 101 <RMT>
Example: none
FETCh[:SCALar]:CURRent:HARMonic:ARRay? <CRD>{<,><NR1>}
MEASure[:SCALar]:CURRent:HARMonic:ARRay? <CRD>{<,><NR1>}
Type: Channel-Specific.
Description: These queries return the amplitude of all the harmonic order. The
return could be -3 or <NR2>. -3: Invalid data when OCR、OCP
occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:CURRent:HARMonic:ARRay?<space><CRD><PMT>,
FETCh:CURRent:HARMonic:ARRay?<space><CRD>,<NR1><PMT
>,
MEASure:CURRent:HARMonic:ARRay?<space><CRD><PMT>,
MEASure:CURRent:HARMonic:ARRay?<space><CRD>,<NR1><P
MT>
Query parameters: <CRD>, VALUE∣PERCENT
<NR1>, 1 ~ 4: Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:CURRENT:HARMONIC:ARRAY<space><NR2>,
<NR2>,<NR2>,...up to 101 <RMT>
Header off: <NR2>,<NR2>,<NR2>,...up to 101 <RMT>
Separator 0: <NR2>,<NR2>,<NR2>,...up to 101 <RMT>
Separator 1: <NR2>;<NR2>;<NR2>;...up to 101 <RMT>
Example: none
FETCh[:SCALar]:SIGMa:POWer:REAL?
MEASure[:SCALar]:SIGMa:POWer:REAL?
Type: Three-Phase.
Description: These queries return the true power. The return could be -1, -2, -3
and <NR2>. -1: The first time integrated calculation is not complete
yet.
-2: RCE represents “range change error” when integration process
is executing.
-3: Invalid data when OVR、OCR、OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:SIGMa:POWer:REAL?<PMT>,
MEASure:SIGMa:POWer:REAL?<PMT>,
Query parameters: none
Return parameters: <NR2>
Header on: :FETCH:SIGMA:POWER:REAL<space><NR2><RMT>