Digital Power Meter 66203/66204 User’s Manual
E-6
The test result of power measurement:
66203/66204 Range Voltage: 600V Range Current: 20A Range
Fluke 5500 Output
Parameter Max. Spec.
Measured
Result
Min. Spec.
480Vrms
16Arms
60Hz
PF = 1
W 7.6996k 7.6604k
PF 1 0.9981
66203/66204 Range Voltage: 300V Range Current: 5A Range
Fluke 5500 Output
Parameter Max. Spec.
Measured
Result
Min. Spec.
240Vrms
4Arms
60Hz
PF = 1
W 962.46 957.54
PF 1 0.9981
66203/66204 Range Voltage: 150V Range Current: 2A Range
Fluke 5500 Output
Parameter Max. Spec.
Measured
Result
Min. Spec.
120Vrms
1.6Arms
60Hz
PF = 1
W 192.49 191.51
PF 1 0.9981
66203/66204 Range Voltage: 60V Range Current: 0.5A Range
Fluke 5500 Output
Parameter Max. Spec.
Measured
Result
Min. Spec.
48Vrms
0.4Arms
60Hz
PF = 1
W 19.249 19.151
PF 1 0.9981
66203/66204 Range Voltage: 30V Range Current: 0.2A Range
Fluke 5500 Output
Parameter Max. Spec.
Measured
Result
Min. Spec.
24Vrms
160mArms
60Hz
PF = 1
W 3.8498 3.8302
PF 1 0.9981
66203/66204 Range Voltage: 15V Range Current: 0.05A Range
Fluke 5500 Output
Parameter Max. Spec.
Measured
Result
Min. Spec.
12Vrms
40mArms
60Hz
PF = 1
W 481.23m 478.77m
PF 1 0.9981