Using Remote Control
5-9
*IDN?
Description: This command queries manufacturer's name, model name, serial
number and firmware version.
Setting syntax: none
Setting parameters: none
Setting example: none
Query syntax: *IDN?<PMT>
Return parameters: <SRD>, “Manufacturer,Model name,Serial number,F/W version,
FPGA version,PCB version”
Information Example
Manufacturer Chroma ATE
Model name 66204
Serial number 66204A000066
F/W version 1.00
FPGA version 1.00
PCB version 1.00
Header on: <SRD><RMT>
Header off: <SRD><RMT>
Query example: none
*RST
Description: This command performs device initial setting.
Setting syntax: *RST<PMT>
Setting parameters: none
Setting example: none
Query syntax: none
Return parameters: none
Query example: none
*SRE
Description: This command sets the service request enable register (SRER).
Setting syntax: *SRE<space><NR1><PMT>
Setting parameters: <NR1>, 0 ~ 255
Setting example: none
Query syntax: *SRE?<PMT>
Return parameters: <NR1>, 0 ~ 255
Header on: *SRE<space><NR1><RMT>
Header off: <NR1><RMT>
Query example: none
*STB?
Description: This command queries the status byte register.
Bit configuration of Status Byte Register
Bit Position 7 6 5 4 3 2 1 0
Condition --- MSS ESB MAV QES CSUM --- ---
ESB = event status byte summary
QES = questionable status summary

Table of Contents